Description
The Handy Press sphygmomanometer analyzer is equipped with a sequential program of 8 different tests recommended by INMETRO ordinances number 153/2005 and 96/2008 and standards NIE-DIMEL006:May/2008 and NIE-DIMEL-097:April/2011 INMETRO – with uncertainty less than 0.8mmHg.
It was specially developed for calibration and evaluation of sphygmomanometers, where it presents specific routines for determining the maximum measurement error, hysteresis error, air leakage and system deflation times.
The Handy Press sphygmomanometer analyzer has intuitive and easy-to-use software. This differential helps the technician in the execution of the chosen test, and, at the end of the last reading, the test report is issued automatically.
The Handy Press enables a sequence of basic tests and 8 more sequences of tests recommended by INMETRO ordinances 153/2005 and 96/2008 and standards NIE-DIMEL-006:May/2008 – NIE-DIMEL-097-2011.
The Handy Press sphygmomanometer analyzer is pre-configured with the following test sequences:
- Air Leakage test (Standard NIE_DIMEL -006 and NIE-DIMEL – 097;
- Rapid deflation test (INMETRO Ordinance No. 96 of March 2008);
- Deflation Valve test;
- Mode 1 test – Rapid Deflation with 5 points;
- Mode 2 test – Rapid Deflation with 10 points;
- Mode 3 test – Load Cycle followed by a Discharge Cycle every 40 mmHg (NIE-DIMEL -006 and NIE-DIMEL – 097);
- Mode 4 test – Charge Cycle followed by a Discharge Cycle with 10 acquisition points;
- Mode 5 test – Load Cycle followed by a 10 and 10 mmHg Unload Cycle (Ordinance 153);
- Test Mode 6 – Two Cycles of Loading and Unloading in a row with 5 points (DOQ-CGCRE-014 and DOQ-CGCRE-017);
- Test Mode 7 – Two Cycles of Loading and Unloading at 40 mmHg in succession (Standard NIE_DIMEL -006 and NIE-DIMEL – 097 + DOQ-CGCRE-014 and DOQ-CGCRE-017);
- Test Mode 8 – Two Cycles of Loading and Unloading followed by 10 points (DOQ-CGCRE-014 and DOQ-CGCRE-017).